Amersham model 6711 Dosimetry Datasets

Note: The following tables are adapted from Rivard et al., Med. Phys. 31, 633-674 (2004).

Table I. NIST standard WAFAC calibration dates for air kerma strength for each manufacturer, and dose rate constant values. Note that for a given source type, the % change in Λ from the 1999 value is different for different seed models.

Manufacturer Source Type NIST date used by ADCL
and NIST as standard
consensusΛ[cGy -h·-1·U-1] % difference in Λ
from 1999 value
Amersham model 6711 125I April 15, 1998 0.965 N/A

Table II.Consensus g(r) values

r[cm] 0.10 0.15 0.25 0.50 0.75 1.00 1.50 2.00
Line source
approximation L=3.0 mm
1.055 1.078 1.082 1.071 1.042 1.000 0.908 0.814
Point source
approximation
0.696 0.853 0.982 1.048 1.036 1.000 0.912 0.819

 

r[cm] 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00
Line source
approximation L=3.0 mm
0.632 0.496 0.364 0.270 0.199 0.148 0.109 0.0803
Point source
approximation
0.636 0.499 0.367 0.272 0.200 0.149 0.110 0.0810

Interpolated data are boldface, and italicized data are non-consensus data obtained from candidate datasets.

Table III. F(r,θ)

Polar angleθ (degrees)

    0 5 10 20 30 40 50 60 70 80 Φan(r)
r [cm] 0.5 0.333 0.400 0.519 0.716 0.846 0.926 0.972 0.991 0.996 1.000 0.973
  1 0.370 0.429 0.537 0.705 0.834 0.925 0.972 0.991 0.996 1.000 0.944
  2 0.442 0.497 0.580 0.727 0.842 0.926 0.970 0.987 0.996 1.000 0.941
  3 0.488 0.535 0.609 0.743 0.846 0.926 0.969 0.987 0.995 0.999 0.942
  4 0.520 0.561 0.630 0.752 0.848 0.928 0.969 0.987 0.995 0.999 0.943
  5 0.550 0.587 0.645 0.760 0.852 0.928 0.969 0.987 0.995 0.999 0.944

Table IV. Dose rates (cGy · h-1 · U-1) as a function of distance for Amersham model 6711 brachytherapy sources using the 1-D dosimetry formalism of Eq. (11) with interpolation for gL(r) and fan(r).

r (cm) 0.5 1.0 1.5 2.0 3.0 4.0 5.0 6.0 7.0
  3.937 0.911 0.368 0.186 0.0643 0.0284 0.0134 0.00688 0.00373